How It Works
EXPlasma Z7

Low-Temperature Sterilization for 
Veterinary Instruments

The EXPlasma Z7 is a compact hydrogen peroxide sterilizer developed for compatible veterinary instruments that may be sensitive to heat and moisture.

Explore the automated cycle, Flash and Standard modes, manufacturer-listed lumen configurations, material compatibility, and essential load-preparation requirements.

 

 

How the EXPlasma Z7 Process Works

01

Prepare the Load

Thoroughly clean, rinse, inspect, and completely dry compatible instruments before packaging.

02

Package & Load

Place instruments in compatible Tyvek pouches and arrange the load according to the EXPlasma Z7 instructions for use.

03

Select the Cycle

Use Flash for compatible surface loads. Use Standard for manufacturer-listed tubular and lumen configurations.

04

Complete the Cycle

Start the selected cycle. Remove the load only after the Z7 indicates successful cycle completion.

Cycle Options

Flash or Standard: Choose the Appropriate Cycle

The EXPlasma Z7 provides two programmed cycles for different compatible loads. Select the cycle according to the instrument type, the Z7 instructions for use, and the instrument manufacturer’s validated reprocessing guidance.

Flash Cycle

Approximately 10 minutes. Intended for compatible instruments requiring surface sterilization. Do not use Flash mode for tubular or lumen devices.

Standard Cycle

Approximately 20 minutes. Used for compatible instruments and all manufacturer-listed tubular and lumen configurations.

Important Lumen Guidance

Use Standard mode only. Thoroughly clean and completely dry each lumen device, seal it individually in a compatible Tyvek pouch, and process no more than four lumens per load.

Instrument Compatibility

Process Only Compatible Veterinary Instruments

The EXPlasma Z7 processes compatible metal and nonmetal instruments at low temperature. It may be appropriate for heat- and moisture-sensitive devices when hydrogen peroxide processing is permitted by the instrument manufacturer.

Compatibility depends on the instrument’s materials, construction, coatings, adhesives, internal geometry, and packaging. Always follow the current EXPlasma Z7 VET user manual and the instrument manufacturer’s reprocessing instructions. Do not assume compatibility based only on a material name.

 

 

Discover What the EXPlasma Z7 Can Bring to Your Practice

Contact JK Medical to discuss instrument compatibility, veterinary workflow, product availability, pricing, and distribution opportunities.

© 2026 JK Medical, Inc. All rights reserved.

Information icon

We need your consent to load the translations

We use a third-party service to translate the website content that may collect data about your activity. Please review the details in the privacy policy and accept the service to view the translations.